Our website features a wide range of standard power conversion and backup products, but we also have the capacity and capability to design and manufacture custom solutions to meet your specific application needs. Contact our team of experienced engineers today to discuss your project. Dismiss
MAIN OUTPUT SPECIFICATIONS | ||
Output power (continuous) (5 s transient) | 85-132/170-264 Vrms 85-132/170-264 Vrms 132-170 Vrms | 1400 W 1800 W |
Nominal DC output voltage (at 800W) (Semi-regulated) | 12 Vout 15 Vout 24 Vout 28 Vout 48 Vout | 12.4 V 15.6 V 25 V 29 V 50 V |
Efficiency | 12 Vout, 115 Vrms, 1400 W 24 Vout, 115 Vrms, 1400 W 48 Vout, 115 Vrms, 1400 W 12 Vout, 230 Vrms, 1400 W 24 Vout, 230 Vrms, 1400 W 48 Vout, 230 Vrms, 1400 W | 87% typ. 89% typ. 90% typ. 89.5% typ. 90.5% typ. 91.5% typ. |
Hold-up time (to -20%) | 12 / 15 Vout 24 / 28 / 48 Vout | 10 ms @ 1400 W 12 ms @ 1400 W |
Maximum load capacitance | 12 Vout 15 Vout 24 Vout 28 Vout 48 Vout | 48,000 μF 40,500 μF 24,000 μF 19,200 μF 6,000 μF |
Output ripple voltage | Switching frequency (20 MHz BW) Twice line frequency (at 800W) | 0.5% p-p 5.0% p-p |
Turn-on delay | 2 s max. | |
Transient response | Iout steps from 50-75% At 0.2 A/μs | 3% typ / 6% max. dev. 100 ms recovery |
Overvoltage protection | Cyclic restart | 110-120% |
Short circuit protection | Cyclic operation | 115% rated Iout |
Total regulation | Over line, load and temperature | ±6.0% |
Auxillary Output | Always on (See Note 1) | 5 V @ 100 mA |
Thermal protection | Automatic recovery | +125 ºC (PCB Temp) |
REMOTE_ENABLE | Input Low Voltage Input High Voltage | 0.45 V (max) 4.15 V (min) |
INPUT SPECIFICATIONS | ||
AC input voltage | Universal range | AC input voltage Universal range 85-264 Vrms |
Input frequency | Input frequency 47-63 Hz | |
Input current | 115 Vrms @ 1400 W 230 Vrms @ 1400 W | 14 Arms 7 Arms |
Power factor | >0.98 | |
Input surge | 264 Vrms (cold start) | 50 A max. |
Internal input fuses | Both AC lines | 20 A |
GENERAL SPECIFICATIONS | ||
Fundamental ripple freq. | Fundamental ripple freq. Input 500 kHz | Fundamental ripple freq. Input 500 kHz |
Output 250 kHz | Output 250 kHz | |
Audible noise Weight ( AQ0800xxxxGC ) ( AQ1100xxxxGC ) ( AQ1400xxxxGC ) | Fan speed varies with temp. | 45 dBA @ 1 m max. 998 g (35.2 oz) 1179 g (41.6 oz) 1179 g (41.6 oz) |
MTBF | MIL-217 Demonstrated | 533 kHours TBD kHours |
ISOLATION SPECIFICATIONS | ||
Isolation voltage | Input to output Input to ground Output to ground (BF & CF) Output to ground (CFD) | 4000 Vrms 1500 Vrms 1500 Vrms 5000 Vpulse |
Insulation resistance | Output to ground | 10 MΩ min. |
ENVIRONMENTAL CHARACTERISTICS | ||
Thermal performance | Operating ambient (see Figure 11) Non-operating ambient | 0 ºC to +70 ºC -40 ºC to +85 ºC |
Relative humidity | Non-condensing | 5-95% RH |
Altitude | Operating Non-operating | 10,000 ft max. 30,000 ft max. |
Random vibration | 5-500 Hz | 0.03 g2/Hz |
Shock | Half-sine, 10 ms, 3 axes | 20 g peak |
EMC CHARACTERISTICS | ||
Conducted emissions | EN55011 and EN55022, FCC part15 | Level B |
Line frequency harmonics | EN61000-3-2 | Class A |
Voltage fluctuations | EN61000-3-3 | Clause 5b |
ESD air See following details | EN61000-4-2 | Level 4, +/-15kV Perf Criteria A, AC Input Connections Perf Criteria B, DC Output Connections |
ESD contact | EN61000-4-2 | Level 4, +/-8kV Perf Criteria A; HCP, VCP, Case |
Radiated immunity | EN61000-4-3 | Level 3, 10V/m IEC60601-1-2 Ed.4 Table 4 28 V/m IEC60601-1-2 Ed.4 Table 9 Perf Criteria A |
Fast transients | EN61000-4-4 | Level 3, 100KHz rep, AC input leads IEC60601-1-2 Ed.4 Table 5 Perf Criteria A |
Line surge immunity | EN61000-4-5 | Level 3 Perf Criteria B |
Perf Criteria B | Perf Criteria B | |
Conducted immunity | EN61000-4-6 | Level 3 Perf Criteria A |
Power freq. mag. Field | EN61000-4-8 | 30 A/m IEC60601-1-2 Ed.4 Table 4 Perf Criteria A |
Voltage dip immunity See following details | EN61000-4-11 | 0% Ut; 0.5 cycle 45° increments 0% Ut; 1 cycle 70% Ut; 0.5s IEC60601-1-2 Ed.4 Table 5 Perf Criteria A, Load Dependent |
Voltage interruptions | EN61000-4-11 | 0% Ut; 5s IEC60601-1-2 Ed.4 Table 5 Perf Criteria B |
NOTES: 1. Derate 1 mA per °C above 50 °C ambient temperature. 2. Leakage currents see following table. |
Selecting the right power supply for your application is not a small task. Let our technical sales engineers guide you through the process of choosing the right solution for your needs.
Below some parameters usually required to size and select the right power supply:
Contact our team of experienced engineers to discuss your unique power conversion and backup needs. We offer a wide range of highly reliable components that can be integrated and customized to create a solution that meets or exceeds your requirements.
For power automation systems please contact us at power.systems.automation@heliosps.com.au
© Copyright Helios Power Solutions 2024