The ECP130 series is designed to minimize the no load power consumption and maximize efficiency to facilitate equipment design to meet the latest environmental legislation. Approved for medical and ITE applications, this range of single output AC-DC power supplies are packaged in a low profile 1.1″ height with a foot print of just 2″ by 3″. The ECP130 provides up to 130W force-cooled or 100W convection-cooled leading to very high power densities of 19.7 W/in3 or 15.1 W/in3 respectively.

The power supply contains two fuses and low leakage currents as required by medical applications and is safety approved to operate in a 80 °C ambient. The low profile and safety approvals covering ITE and medical standards along with conducted emissions to EN55011/22 level B allow the versatile ECP130 series to be used in a vast range of applications.

SERIES TABLE

Model InputOutputPower
ECP130PS1285 to 264 VAC12 V / 10.83 A130 W
ECP130PS1585 to 264 VAC15 V / 8.66 A130 W
ECP130PS1885 to 264 VAC18 V / 7.22 A130 W
ECP130PS2485 to 264 VAC24 V / 5.41 A130 W
ECP130PS2885 to 264 VAC28 V / 4.68 A130 W
ECP130PS3685 to 264 VAC36 V / 3.61 A130 W
ECP130PS4885 to 264 VAC48 V / 2.70 A130 W

SPECIFICATIONS TABLE

Input Voltage:85 to 264 VAC. Derate load from 100% at 90 VAC to 85% at 85 VAC
No Load Input Power<0.5 W
Harmonic Currents:EN61000-3-2, class A
Efficiency:95% typical
Operating Temp:-30 °C to +80 °C. See datasheet for derating.
Isolation:4000 VAC Input to Output 2 x MOPP
1500 VAC Input to Ground 1 x MOPP
1500 VAC Output to Ground 1 x MOPP
Worldwide Safety Approvals:IEC60950-1:2005, IEC60601-1 Ed 3.1 Including Risk Management
European Safety Approvals:EN60950-1, EN60601-1, CE Mark LVD
US Safety Approvals:UL60950-1, CSA 22.2 No.60950-1-11 Ed 2, ANSI/AAMI ES60601-1: & CSA C22.2 No.6061-1:08
EMC:EN55011 class B conducted
EN55011 class A radiated (class B – see datasheet)

Ask for Assistance

Selecting the right power supply for your application is not a small task. Let our technical sales engineers guide you through the process of choosing the right solution for your needs.

Below some parameters usually required to size and select the right power supply:

  • Power & control requirements
  • Output & Input Voltage / Multioutput Voltage / Variable Voltage
  • Redundancy level
  • Mounting
  • Inrush Current
  • Environmental specifications (Temperature, altitude, vibration, shock)
  • Specific Industries Standards (Railway, Military, Medical)
  • Terminals

Contact Our Team

Contact our team of experienced engineers to discuss your unique power conversion and backup needs. We offer a wide range of highly reliable components that can be integrated and customized to create a solution that meets or exceeds your requirements.

For power automation systems please contact us at power.systems.automation@heliosps.com.au